¢Â Fluke Documenting Process Calibrator |
|
|
|
|
|
|
|
|
|
 |
|
|
|
Á¦¸ñ ¾øÀ½
Measurement Accuracy |
Voltage DC |
110.000 mV |
0.025%+0.015% * |
1.10000 V |
0.025%+0.005% * |
11.0000 V |
0.025%+0.005% * |
110.000 V |
0.05%+0.005% * |
300.00 V |
0.05%+0.005% * |
Voltage AC |
20 to 40 Hz |
2% + 10 |
40 to 500 Hz |
0.5% + 5 |
500 to 1 kHz |
2% + 10 |
1 kHz to 5 kHz |
10% + 20 |
Ranges: |
1.1000, 11.000, 110.00, 300V |
Current DC |
30.000 mA |
0.01% + 0.015% * |
110.00 mA |
0.01% + 0.015% * |
Resistance |
11.000 ¥Ø |
0.05% + 50 m¥Ø |
110.00 ¥Ø |
0.05% + 50 m¥Ø |
1.1000 k¥Ø |
0.05% + 500 m¥Ø |
11.000 k¥Ø |
0.1% + 10 ¥Ø |
Frequency |
1.00 to 109.99 Hz |
0.05 Hz |
110.0 to 1099.9 Hz |
0.5 Hz |
1.100 to 10.999 kHz |
5 Hz |
11.00 to 50.00 kHz |
50 Hz |
Pressure |
Accuracy from 0.025% of range using any of 29 pressure modules. Modules available for differential, gauge, vacuum, absolute, dual and high pressure. (for detailed specifications refer to pressure modules in options and accessories) |
Note |
* (% of reading + % of full scale) |
Source Accuracy |
Voltage DC |
110.000 mV |
0.01%+0.005% * |
1.10000 V |
0.01%+0.005% * |
15.0000 V |
0.01%+0.005% * |
Current DC |
22.000 mA (Source) |
0.01%+ 0.015% * |
22.000 mA (Simulate) |
0.02% + 0.03% * |
Resistance |
11.000 ¥Ø |
0.01% + 20 m¥Ø |
110.00 ¥Ø |
0.01% + 40 m¥Ø |
1.1000 k¥Ø |
0.02% + 500 m¥Ø |
11.000 k¥Ø |
0.03% + 5¥Ø |
Frequency |
0.00 to 10.99 Hz |
0.01 Hz |
11.00 to 109.99 Hz |
0.1 Hz |
110.0 to 1099.9 Hz |
0.1 Hz |
1100 to 21999 Hz |
2 Hz |
22.000 to 50.000 kHz |
5 Hz |
Note |
* (% of reading + % of full scale) |
RTDs and Thermocouples |
Measure Accuracy: |
10 ¥Ø Cu (427): |
2 ¡ÆC |
100 ¥Ø Pt (3916): |
0.3 ¡ÆC |
100 ¥Ø Pt (3926): |
0.3 ¡ÆC |
100 ¥Ø Pt (385): |
0.3 ¡ÆC |
200 ¥Ø Pt (385): |
0.3 ¡ÆC |
500 ¥Ø Pt (385): |
0.3 ¡ÆC |
1000 ¥Ø Pt (385): |
0.3 ¡ÆC |
120 ¥Ø Ni (672): |
0.3 ¡ÆC |
note: |
For 2-and 3-wire measurement, add 0.4¡ÆC |
E: |
0.3 ¡ÆC |
N: |
0.5 ¡ÆC |
J: |
0.3 ¡ÆC |
L: |
0.3 ¡ÆC |
K: |
0.3 ¡ÆC |
T: |
0.3 ¡ÆC |
U: |
0.3 ¡ÆC |
B: |
0.9 ¡ÆC |
R: |
1.0 ¡ÆC |
S: |
0.9 ¡ÆC |
C: |
0.6 ¡ÆC |
BP: |
1.2 ¨¬C |
XK: |
0.4 ¨¬C |
note: |
Accuracy with external cold junction, for internal junction add 0.2 ¡ÆC |
Source Accuracy: |
10 ¥Ø Cu (427): |
1 ¡ÆC |
100 ¥Ø Pt (3916): |
0.1 ¡ÆC |
100 ¥Ø Pt (3926): |
0.1 ¡ÆC |
100 ¥Ø Pt (385): |
0.1 ¡ÆC |
200 ¥Ø Pt (385): |
0.1 ¡ÆC |
500 ¥Ø Pt (385): |
0.1 ¡ÆC |
1000 ¥Ø Pt (385): |
0.1 ¡ÆC |
120 ¥Ø Ni (672): |
0.1 ¡ÆC |
note: |
For 2-and 3-wire simulation, add 0.4 ¡ÆC |
E: |
0.2 ¡ÆC |
N: |
0.3 ¡ÆC |
J: |
0.2 ¡ÆC |
L: |
0.2 ¡ÆC |
K: |
0.3 ¡ÆC |
T: |
0.3 ¡ÆC |
U: |
0.3 ¡ÆC |
B: |
0.8 ¡ÆC |
R: |
0.9 ¡ÆC |
S: |
0.9 ¡ÆC |
C: |
0.6 ¡ÆC |
BP: |
0.5 ¨¬C |
XK: |
0.4 ¨¬C |
note: |