Milliamp
Source |
|
Ranges: |
0.000 to 24.000mA; -25.00 to 125.00% of 4 to
20mA |
Accuracy: |
¡¾(0.05% of 24 mA Span + 0.01 mA) = 0.02mA |
Typical Drive Capability: |
1200 Ohms @ 20.00 mA |
Compliance Voltage: |
Nominal 25 V @ 20 mA |
Power &
Measure 2-wire Transmitters |
|
Ranges: |
0.000 to 24.000mA; -25.00 to 125.00% of 4 to
20mA |
Accuracy: |
¡¾(0.05% of 24 mA Span + 0.01 mA) = 0.02mA |
Output Current: |
Up to 24.00 mA |
Typical Drive Capability: |
1200 Ohms @ 20.00 mA |
Compliance Voltage: |
Nominal 25 VDC @ 20mA |
2-wire
Transmitter Simulator |
|
Ranges: |
1.000 to 24.000mA; -18.8 to 125.00% of 4 to
20mA |
Accuracy: |
¡¾(0.05% of 24 mA Span + 0.01 mA) = 0.02mA |
Loop Voltage Limits: |
Minimum: 3 VDC Maximum: 45 VDC |
Overload Protection: |
Current limited to 25 mA nominal |
Voltage Burden: |
0.9V at mA, 1.2V at 20mA, 1.9V at 24mA |
Milliamps
Read |
|
Ranges: |
0.000 to 24.000mA; -18.8 to 125.00% of 4-20mA |
Accuracy: |
¡¾(0.05% of 24 mA Span + 0.01 mA) = 0.02mA |
Overload Protection: |
Current limited to 25 mA nominal |
Voltage Burden: |
0.9V at 4 mA, 1.2V at 20 mA, 1.4V at 24 mA |
DC Voltage
Source |
|
Ranges: |
0.000 to 110.000mV 0.00 to 10.25V |
Accuracy: |
¡¾(0.05% of 110 mV+ 0.01mV) = ¡¾0.07 mV ¡¾(0.05% of 10.25
V + 0.01V) = ¡¾0.02V |
Source Current: |
>20mA |
Sink Current: |
>20mA |
Output Impedance: |
Greater than 0.3 Ohms |
Short Circuit Duration: |
Infinite |
Measure AC
Volts |
|
Ranges: |
0.0 to 250.0V True RMS |
Accuracy: |
From 10 to 250 VAC ¡¾(2% of 250V + 0.1 VAC) =
¡¾5.1VAC |
Maximum Crest Factor: |
Greater than 3 |
Frequency Range: |
45 to 800 Hz |
General |
|
Typical 90 day Accuracy: |
¡¾(0.025% of Full Scale + 1 LSD) |
1 Year Accuracy: |
¡¾(0.05% of Full Scale + 1 LSD) |
Warm up Time: |
10 Seconds to specified accuracy, 2 minutes to maximum
accuracy |
Temperature Effect: |
¡¾0.01%/¡ÆC based on 23¡ÆC ¡¾25¡ÆC |
Batteries: |
Six "AA" (R6) batteries (Alkaline supplied and
recommended) |
Battery Life: |
Milliamp Source and 2-wire Modes: Nominal 50 hours
at 12 mA, 20 hours at 20 mA with 250 ohm load Other Functions: Nominal
50 hours |
Low Battery Indication: |
"BAT" indication on the display, approximately 4 hours
left |
Overload Protection: |
Three fuses, 250 mA |
Noise: |
=1 LSD at frequencies less than 10 Hz |
Normal Mode Rejection Ratio: |
50dB @ 50/60 Hz |
Operating Temperature Range: |
-5 to +130 ¡ÆF (-20 to +55¡ÆC) |
Storage Temperature Range: |
-13 to +130 ¡ÆF (-25 to +55¡ÆC) |
Relative Humidity: |
10 to 90%, non-condensing for 24 hours |
Dimensions: |
158.1mm (6.23 in) x 83.1mm (3.27 in) x 49.3mm (1.94
in) |
Weight: |
1lb 5oz (0.6kg) |